Technical Bulletin: AV-9900 Time
Change
UNIT: AV-9900
HTST Recorder / Controller
Revision: 1.0 | Doc
#: TS-3000
The following
procedure covers time change on the AV-9900 HTST.
- The unit may have
the Health Authority seal left in place, with the unit in RUN mode
- The chart must
NOT be in motion - a STOP CHART must be performed
- Please note that
the stop chart function will cause a system divert if in forward
flow
- Press [SCROLL]
until you reach [CONFIGURATION]
- Press [DOWN] arrow
key
- Press [SCROLL]
until you reach [INSTRUMENT SETTINGS]
- Press [DOWN] key
- Press [SCROLL]
key until you reach [TIME]
- Press [MOD] key
- Use [UP] and [DOWN]
arrows to change HOUR
- Press [ENTER]
- Use [UP] and [DOWN]
arrows to change MINUTE
- Press [ENTER]
- Use [UP] and [DOWN]
arrows to change AM / PM
- Press [ENTER]
- Time change is
now complete
- Press [RESET] to
return to main menu