

Technical Bulletin - SL/SX Level Sensor Troubleshooting
UNIT: Model SL and SX Liquid Level Transmitter
Revision: 1.0 | Doc #: TS-3001
Symptoms, and their associatd corrective actions have been provided for the purpose of field troubleshooting the SL/SX Liquid Level Transmitters.
Note: As many sensors provide feedback to cooling systems, agitator controls or lockout controls be sure to properly secure all associated systems prior to performing troubleshooting.
CLICK HERE to view the complete service manual
NO CURRENT OUTPUT
Loop may be broken - Measure voltage across LOOP(+) and LOOP(-) screws. Result should be between 12-40 VDC.
Sensor circuitry damaged - Move LOOP(+) wire to TEST(-) screw. If output loop now works, internal sensor circuit has been damaged. Contact Anderson.
OUTPUT CURRENT STUCK BETWEEN 4 AND 20 mA
Sensor must be in RUN mode - be sure sensor is NOT in CURRENT CAL or FIELD CAL switch positions.
Check diaphragm for damage - dent or ding may cause output shifts beyond adjustment. Contact Anderson.
Verify that CURRENT CAL value is correct - Section 3.3 of Service Manual.
CANNOT SET TO 4.00 mA (+/- .04 mA) USING ZERO SWITCH
Verify that CURRENT CAL value is correct - Section 3.3 of Service Manual.
If current value is greater than 4.00 mA, sensor is damaged. Contact Anderson.
If current is less than 4.00 mA, follow checks in "No Current Output" section above.
SENSOR OUTPUT SIGNAL NOT STABLE - DRIFTS
Verify that CURRENT CAL value is correct - Section 3.3 of Service Manual.
Check diaphragm for damage - dent or ding may cause output shifts beyond adjustment. Contact Anderson.
Check for signs of moisture / water in housing. If found, check cable entry. Must be sealed to prevent moisture entry. Cap must be tightened completely to engage gasket. Removing sensor from service - place in low humidity area with cap removed. Sensor may function correctly when dry.
CURRENT OUTPUT VALUE INCORRECT
Follow section 3.3 of Service Manual - Verifying Sensor Calibration. Sensor may have been adjusted improperly with product in vessel.
If calibration does not match factory setting, calibrate as shown in Section 3.5 of Service Manual.
OUTPUT CURRENT JUMPS TO VALUE GREATER THAN 23 mA
Follow section 3.3 of Service Manual - Verifying Sensor Calibration. Sensor may have been adjusted improperly with product in vessel.
If calibration does
not match factory setting, calibrate as shown in Section 3.5 of Service
Manual.

| Service & Support Material |
| Manuals & Bulletins |
| Recorders |
| Pasteurization |
| Liquid Level |
| Pressure |
| Temperature |
| Instrumentation |
| Analytical/Flow |
| Tech Service Contacts |
| Field Service Training Course |