

![]() | |
| |
| My old unit was an SE - This one looks different - is it compatable? | |
| What needs to be done when I install a new sensor? | |
| Can I re-calibrate a sensor in the field? | |
| Are the ZERO and SPAN settings affected by each other? | |
| Can I perform a SPAN change with product in the vessel? | |
| Is there a quick troubleshooting guide? | |
| Display of level indicator goes to zero when the test meter is disconnected, but reads correctly while meter is connected across the Test+ and Test- terminals. | |
| My old unit was an SE - This one looks different - is it compatable? | |
| The SL/SX sensors are an upgrade to the previous SE sensor model. Function is identical as far as both provide a 4-20 mA signal output - wiring is the same. Note that ranges should be reviewed. SL/SX sensors may use different ranges than the SE to cover the same calibration range. | |
| What needs to be done when I install a new sensor? | |
In most cases, sensors are supplied pre-caibrated directly from the factory. Details are as follows: 1. Be sure
to follow tag to ensure watertight cable entry. For complete
service manual - Click to See (Pg 11 shows details) | |
| Can I re-calibrate a sensor in the field? | |
| There are
four ranges of sensor - desired new calibration MUST be within
range of that model. Instructions are provided in the Service
Manual - Click to See. | |
| Are the ZERO and SPAN settings affected by each other? | |
| No - Zero and Span settings are non-interactive. When a Zero is performed, it is not necessary to perform a check of the Span. | |
| Can I perform a SPAN change with product in the vessel? | |
| Yes - Changes to the span may be made using the on-board Field Cal feature of the sensor, even with product in the vessel. CAUTION - Performing this operation will affect the current output of the unit - be sure to verify all automated controls first. | |
| Is there a quick troubleshooting guide? | |
| Yes - Technical Bulletin TS-3001 shows quick troubleshooting tips. - Click to See | |
| Display of level indicator goes to zero when the test meter is disconnected, but reads correctly while meter is connected across the Test+ and Test- terminals. | |
| The sensor may have been temporarily miswired, or the meter leads improperly located, resulting in a component failure within the sensor. A wire jumper may be connected between the Test+ and Test- terminals to bypass the failed component. The Test terminals will no longer be useable. | |

| Service & Support Material |
| Manuals & Bulletins |
| Recorders |
| Pasteurization |
| Liquid Level |
| Pressure |
| Temperature |
| Instrumentation |
| Analytical/Flow |
| Tech Service Contacts |
| Field Service Training Course |